Asbestos in Water Testing Overview & FAQ

Background

In 1974 the EPA passed the Safe Drinking Water Act (SDWA) in an effort to determine the level of contamination in drinking water to which no adverse health effects are likely to occur. The first Asbestos in Water method, EPA 100.1, was written in 1983 as the “Analytical Method for Determination of Asbestos Fibers in Water” (EPA-600/4-83-043). In 1992, the EPA established the first maximum contaminant level and added to the Federal Register that the analysis of asbestos in water should be conducted using Transmission Electron Microscopy (TEM). It was then that the EPA 100.1 method did not meet the needs of this new 1992 regulation, so the EPA issued a memorandum to clarify how the EPA 100.1 standard could be used for compliance. A new method EPA 100.2 was issued in 1994 (“Determination of Asbestos Structures Over 10µm in Length in Drinking Water, EPA/600/R-94/134) and was written specifically to respond to the asbestos drinking water standard. It also incorporated quality assurance objectives developed under the Asbestos Hazard Emergency Response Act (AHERA). Asbestos was added as a contaminant to the Safe Water Drinking Act in 1992, and was added to the National Primary Drinking Water Regulations (NPDWR) in 1995. These regulations are legally enforceable primary standards that apply to public water systems only. The EPA has set a maximum contaminant level of 7 million fibers (>10 micron in length), per liter (MFL). The EPA periodically reviews this act to determine if the MCL still protects human health. States and other local agencies may set more stringent drinking water MCL’s for asbestos than the EPA, so it is always best to check with your local authorities prior to interpreting your laboratory results.

Test Method: EPA 100.2 - Asbestos in Drinking Water (EMSL recommended method)

The EPA 100.2 method is recommended for the determination of the identification and quantification of asbestos in water (potable and non-potable). Only asbestos fibers greater than 10µm in length are counted in this method, however all fiber sizes >0.5µm may be reported upon request for an additional charge. Although most agencies use the EPA regulatory limit (7MFL asbestos fibers >10µm) it is best to check with your state to ensure the regulatory limit and fiber size requirement. Excessive sediment in a water sample reduces the amount of water that can be filtered to achieve a “readable” grid, so lesser volumes must be filtered, raising the analytical sensitivity. An elevated analytical sensitivity may be reduced by analyzing more filter area, but this is an additional cost.

Test Method: EPA 100.1 - Asbestos in Water

Although still a valid, EPA method EPA 100.1 has widely been replaced by EPA 100.2. EPA 100.1 is for determining the concentration of asbestos fibers in water samples and counts asbestos fibers >0.5 µm in length with aspect ratios of 3:1 or greater. Results are expressed in millions of fibers per liter (MFL). Excessive sediment in a water sample reduces the amount of water that can be filtered to achieve a “readable” grid, so lesser volumes must be filtered, raising the analytical sensitivity. An elevated analytical sensitivity may be reduced by analyzing more filter area, but often at additional cost. EPA 100.1 is only available in select EMSL laboratories. Please check with your sales representative.

Comparison of Methodologies: EPA 100.2 Versus EPA 100.1

Test:EPA 100.1EPA 100.2
Typical Use:Research ProjectRegulatory compliance
Matrices Accepted:Potable / Non-Potable WaterPotable / Non-Potable Water
Sample Collection:Two (2)- 1 liter polyethylene (recommended) or amber glass containers.
Containers are to be analyzed for background levels of asbestos or certified clean
prior to sample collection. EMSL Product Number 8714233. $5.00 Each
Ozone/UV treatment
(limited labs):
Samples must be ozone treated and
filtered within 48 hours of collection.
Samples must be filtered within 48
hours of collection. Ozone treatment
may be performed on samples past
the 48 hour hold and/or samples with
organic particulate.
Preservative/ Ozone Treatment:EMSL Analytical does not use mercuric
chloride preservative. Samples free
of organic material must be filtered
within 48 hours of collection or ozone
treatment is required. Samples must
be received at a temperature of 4°±2°C
or within 24 hours of collection on ice.
Samples must not be frozen.
This method does not use mercuric
chloride preservative. Samples free
of organic material must be filtered
within 48 hours of collection or ozone
treatment is required. Samples must
be received at a temperature of 4°±2°C
or within 24 hours of collection on ice.
Samples must not be frozen.
Filter:PolycarbonateMixed Cellulose Ester or Polycarbonate
Filter Size:25 mm or 47 mm25 mm or 47 mm
Filter Pore Size:0.1 µm0.1 to 0.22 µm
Minimum fiber size counted> 0.5 µmTwo options: > 10 µm ; >0.5µm
Minimum aspect ratio3:13:1
Analysis Stopping Rules
Minimum Grid Openings Counted
Number of Structures Counted
Analytical Sensitivity

4
100
not specified

4
100
0.2 MFL
Magnification20,000 X10,000 - 20,000 X
Analytical Sensitivity> 0.1 million fibers per liter (MFL)0.2 million fibers per liter (MFL)
Sample BlanksNot a requirement of either method.

Frequently Asked Questions

Q.  My sample arrived at the lab after the 48-Hour Hold Time, what happens now?
A. Three of our laboratories in the EMSL network (Our Corporate Laboratory in Cinnaminson, NJ; Minneapolis, MN; and our LA Testing laboratory in South Pasadena, California) have Ultra Violet (UV) Light / Ozone treatment capabilities. In order for analysis to be compliant with the EPA method, this treatment must be used when samples are not able to be filtered within (48 hours).

There is an additional preparation charge of $90.00 per sample for UV / Ozone Treatment.

Q. My state / governing agency requires that all asbestos fiber >0.5µm be included in the count?
A. All of our laboratories can perform an EPA 100.2 analysis and include all fibers > 0.5µm. You will need to denote this when you submit the samples. Make this selection on EMSL’s Chain of Custody Document:

EPA Method 100.2 (All fiber sizes ≥ 0.5μm)

Q. My State/ Governing Agency requires that analysis be done using 100.1 method.
A. Due to the requirement for Ozone treatment for this method, only our Cinnaminson, NJ / Minneapolis, MN / and South Pasadena, CA locations can perform this method.

Q. I am submitting non-potable water for analysis. What analysis method should I choose and do I need Ozone treatment?
A. Both of the EPA water methods may be used for non-drinking (non-potable) water samples. The EPA 100.2 is the more widely recognized, and a lower cost. If your samples are groundwater or waste water, it is possible they will require Ozone/UV treatment prior to analysis to remove organic materials which can cause under reporting of asbestos.

Q. What type of water bottle should I utilize?
A. EMSL offers water bottles for both the EPA 100.1 and 100.2 methods. The EPA 100.2 method states that a ‘fiber-free’ bottle should be utilized in the collection of drinking- water samples. EMSL offers these ‘fiber-free’ bottles for $5.00 each – EMSL Product ID: 8714233.

EPA 100.2 Method – Section 8.1:

8.0

SAMPLE COLLECTION, PRESERVATION AND STORAGE

8.1

The sample container will be an unused, pre-cleaned, screw-capped bottle of glass or low density (conventional) polyethylene of at least 1 liter capacity. It is recommended that the use of polypropylene bottles be avoided since problems of particulate being released into water samples have been observed. Before use, the bottles should be rinsed twice by filling approximately one- third full with fiber-free water and shaking vigorously for 30 seconds. After discarding the rinse water, the bottles should then be filled with fiber-free water and treated in an ultrasonic bath for 15 minutes, followed by several rinses with fiber-free water.

Frequently Asked Questions – Compliance Drinking Water Sampling

It is important to note that drinking water samples that are used for compliance to the safe drinking water act must comply with perfect adherence to the method. A water district will not accept a qualifier on the report. There are times where drinking water samples are used for home or building owners where it is acceptable to have a qualifier to the report while satisfying the intent of the results, for example the water arrived at the lab not on ice at a temperature of 7°C, a footnote must be issued in the report stating that the temperature at time of arrival was outside the acceptable limit (2°-6°C), while this would be turned away from a water district it may be completely acceptable result for the client. Below is a list of requirements and typical reports qualifiers we have run across.

SamplingDiscussionQualifier/resolution
Bottles – Both 100.2 and
100.1 require the use of two
(2) 1L pre-cleaned bottles
Only one bottle submittedQualifier required – two precleaned
bottles must be submitted
Samples are not pre-cleanedQualifier required – pre-cleaned bottles
are required
Pre-cleaned from an outside source
were used
Use EMSL Chain of Custody or clearly
and mark that precleaned bottles have
been used on your chain of custody.
Please note you may be asked to show
proof.
Temperature of samples are
below 2 or above 6°C
Samples arrive within 24 hours of
collection time on ice
Acceptable – no disclaimer required
Samples arrive outside of required
temperature
Qualifier required
Particulate LoadingAt times particulate matter may affect
the analytical sensitivity to a point
where it is not possible to achieve
without advanced analytical effort.
Qualifier required if analytical sensitivity
cannot be reached. Note: the sample
maybe able to reach the analytical
sensitivity with an increased analytical
effort (addition cost per grid opening).
Unfortunately, this option is not always
feasible- please contact the laboratory.
Collection ActivitiesIt is imperative to have all required
information available for data entry in
the regulatory system.
Utilizing EMSL Analytical chain of
custody. It is also recommended when
sampling for the first time to check with
the lab to see if they have an account
with the state.
Ozone TreatmentEPA 100.2 samples beyond the 48-hour
hold may be ozone treated
Report will have comment that ozone
treatment was performed but it is
part of the method and therefore not
considered a qualifier and acceptable to
report.

For More Information on Our Lab Testing Services, Pricing, Sampling Guides, or Products: