MoldSnap and Micro-5 Sampling Guide
The Zefon MoldSnap (Product ID 8715304) and the Micro5 MicroCell (Product ID 8715306) were designed to operate at a flow rate of 5 liters per minute for optimal collection efficiency. Cassettes should be stored at 50° – 80° F.
Do not use in temperatures below 37° F!
Sampling Procedures
- Remove pin from the bottom of the MoldSnap or Micro5.
- If using a conventional high volume pump: Simply connect one end of tubing (5 feet or less) to the bottom of the cassette.
- If using a low volume pump such as a GilAir5: Simply attach a 1” piece of tubing to the bottom of the cassette.
- Connect the other end of tubing to a pump pre-calibrated to a flow rate of 5 LPM.
- Remove cap from the top of the cassette.
- Turn pump on and take a sample for an appropriate amount of time depending on environment. (See “Sampling Duration” below.)
- When sampling is completed, replace the pin and cap to the bottom and top of the cassette.
Sampling Duration
- Outdoor / Indoor Clean Environments = 8-10 minutes
- Indoor (Normal Activity) = 5 minutes
- Indoor (Heavy Particulates) = 1-3 minutes
- Inner wall cavity using the Micro5 only (with optional Inner Wall Adapter, Product ID 8715901) = 1-2 minutes