MoldSnap and Micro-5 Sampling Guide

The Zefon MoldSnap (Product ID 8715304) and the Micro5 MicroCell (Product ID 8715306) were designed to operate at a flow rate of 5 liters per minute for optimal collection efficiency. Cassettes should be stored at 50° – 80° F.

Do not use in temperatures below 37° F!

Sampling Procedures

  1. Remove pin from the bottom of the MoldSnap or Micro5.
  2. If using a conventional high volume pump: Simply connect one end of tubing (5 feet or less) to the bottom of the cassette.
  3. If using a low volume pump such as a GilAir5: Simply attach a 1” piece of tubing to the bottom of the cassette.
  4. Connect the other end of tubing to a pump pre-calibrated to a flow rate of 5 LPM.
  5. Remove cap from the top of the cassette.
  6. Turn pump on and take a sample for an appropriate amount of time depending on environment. (See “Sampling Duration” below.)
  7. When sampling is completed, replace the pin and cap to the bottom and top of the cassette.

Sampling Duration

  • Outdoor / Indoor Clean Environments = 8-10 minutes
  • Indoor (Normal Activity) = 5 minutes
  • Indoor (Heavy Particulates) = 1-3 minutes
  • Inner wall cavity using the Micro5 only (with optional Inner Wall Adapter, Product ID 8715901) = 1-2 minutes

For More Information on Our Lab Testing Services, Pricing, Sampling Guides, or Products:

Subscribe to Our Monthly Email Newsletter for Industry Professionals.

EMSL's EASY TO USE TEST KITS

Trusted by Inspectors, Property Managers, Building Owners, Real Estate and Environmental Professionals

Our test kits offer simple, easy-to-follow instructions that allow both the do-it-yourself or the field professional an added testing service as part of an inspection or prevention strategy. All kits include pre-paid postage and a variety of turnaround options including same-day!

*Lab Fees Required Upon Return For Testing